Confirm a maintenance order for a calibration task and verify the equipment's next calibration due date is recalculated correctly based on the configured cycle. Test with a non-standard confirmation date (early or late completion) to confirm the system calculates the next due date from the correct reference point — confirmation date vs originally scheduled date — matching your URS requirement.
Allow a calibration due date to pass without confirming the corresponding maintenance order. Confirm the equipment status changes automatically to the configured overdue/locked status. Document the timing — does the status change exactly on the due date, or is there a grace period configured, and does that match the URS.
With equipment in a "Calibration Overdue" or "Locked" status, attempt to release a process order or confirm an operation that requires this equipment. Confirm the system blocks the action with a clear error message. This is the single most important PM test — it confirms the calibration gate is a system control, not a procedural one. If the system only warns rather than blocks, document this explicitly as it changes the validation conclusion.
Attempt to manually change equipment status from "Locked" to "Available" using a standard user role. Confirm the action either requires elevated authorisation or generates a mandatory audit trail entry with a reason code. Test that a user without the override authorisation profile is blocked from performing this action.
Confirm a maintenance order with measurement results entered. Retrieve the change document and confirm it captures technician ID, confirmation timestamp, and the recorded values. Attempt to edit the confirmation text after saving and confirm a change record is generated — this tests whether retrospective edits to completed calibration records are traceable.