Submit results with maximum expected decimal precision, results at zero or minimum plausible value, and a result transmitted in a non-default unit if the instrument supports it. Confirm the LIMS receives and stores each value exactly as generated, with no truncation and no silent unit conversion error.
Deliberately interrupt the connection (network disconnect, instrument software interrupt) during an active data transfer. Confirm the LIMS either records no data for that transfer or records a clearly flagged incomplete/failed state — not a partial record indistinguishable from a complete, valid result.
Submit the identical result twice in succession. Confirm the system's behaviour (reject second, flag for review, or require manual reconciliation) matches the documented URS requirement, and that any accepted duplicate generates an audit trail entry rather than silently overwriting the original.
Run multiple samples through the instrument in immediate succession. Confirm each result in LIMS is linked to the correct sample ID and test — not offset or mismatched due to timing or queue-handling errors in the interface software.
For bidirectional interfaces, confirm the sample ID and method parameters transmitted from LIMS to the instrument match exactly what was configured at sample login. Test with multiple samples queued simultaneously to confirm no cross-assignment occurs between samples in the same batch.
If the instrument's own software performs calculations before sending results, cross-reference the instrument software's own report against the LIMS-received value for a representative sample set, confirming the transferred value matches what the instrument software actually calculated — not just that a plausible number arrived.